Soft X-Ray Microcharacterization Beamline

Solid State Endstation

Bulk analysis of solid samples, such as pellets and powders can be performed using this endstation. Samples are placed under vacuum, optimizing flux for lower energy edges such as silicon, phosphorus, and sulfur.  This endstation is equipped with fluorescence and total electron yield (TEY) detection.

Capabilities

  • XAFS spectroscopy under vacuum:~5 × 10⁻7 Torr

  • X-Ray Fluorescence Yield (XRF)

  • Total Electron Yield (TEY)

  • X-ray Excited Optical Luminescence (XEOL)

  • Automated determination of sample positions

  • Motorized sample stage

  • Automated venting/pumpdown system

Samples

  • Solid

  • Foil

  • Thin Film

A sample holder can hold up to 40 samples using double sided carbon tape.

Specifications

Source

Bending Magnet

Energy Range (keV)

1.7 - 10

Wavelength (Å)

7.3 - 1.3

Resolution ΔE/E @ E InSb (111)

3.3 × 10⁻⁴

Resolution ΔE/E @ E Si (111)

1.0 × 10⁻⁴

Flux (g/s/0.1%BW) @ 100 mA

> 1.0 × 10¹¹

Spot Size (V × H)

~ 1 mm × 4 mm

Focusing

Torroidal Mirror

Detectors

7-element Si drift, Bruker, Vortex

image-20261009-035517.png
d33268f4-dd35-4528-8f77-919b3ed0344e.png

For more information regarding SXRMB beamline, please refer to the following paper:

Preliminary Commissioning and Performance of the Soft X‐ray Micro‐characterization Beamline at the Canadian Light Source