🟢 Available
The microprobe endstation operates under vacuum and is specifically tailored for applications requiring the challenging medium-energy range of 2-4 keV. This endstation utilizes a pair of Kirkpatrick-Baez (K-B) mirrors to focus the collimated beam to a spot size of 10 µm × 10 µm for mapping experiments. A high-resolution and large-area CCD camera is equipped to obtain an optical image of the sample. Micro-XAS can also be performed on specific spots using the microprobe.
Capabilities
-
Vacuum compatible
-
Sample area size: 18 x 18 mm
-
Micro focusing (0.15×0.15 mrad2, hor.×ver.) optics
-
30 cm working distance
-
Automated venting/pumpdown system
-
Mapping in fluorescence and TEY modes
-
Micro-XANES
Samples
-
Thin Film
-
Solid (i.e., a uniform, thin layer of powder or a surface polished bulk sample)
-
Potential application: imaging of geological and biological samples at ~10 µm
Schematic
Specifications
|
Source |
Bending Magnet |
|---|---|
|
Energy Range (keV) |
1.7 - 10 |
|
Wavelength (Å) |
7.3 - 1.3 |
|
Resolution ΔE/E @ E InSb (111) |
3.3 × 10⁻⁴ |
|
Resolution ΔE/E @ E Si (111) |
1.0 × 10⁻⁴ |
|
Flux (g/s/0.1%BW) @ 100 mA |
1.0 × 10⁹ |
|
Spot Size (V × H) |
~10 µm × 10 µm |
|
Focusing |
K-B Mirror Pair |
|
Mirror characteristic |
20 cm long with a C/Pt bi-layer coating |
|
Detector |
7-element Si drift |
For more information regarding the microprobe endstation at the SXRMB, please refer to the following paper:
Medium-energy microprobe station at the SXRMB of the CLS